Title of article
Analysis of 18F-labelled synthesis products on TLC plates: Comparison of radioactivity scanning, film autoradiography, and a phosphoimaging technique
Author/Authors
Eeva-Liisa K?m?r?inen، نويسنده , , Merja Haaparanta، نويسنده , , Marja Siitari-Kauppi، نويسنده , , Teija Koivula، نويسنده , , Tiina Lipponen، نويسنده , , Olof Solin، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
5
From page
1043
To page
1047
Abstract
We compared radioactivity scanning, film autoradiography, and digital photostimulated luminescence (PSL) autoradiography (phosphoimaging technique) in detection of radioactivity on thin-layer chromatography (TLC) plates. TLC combined with radioactivity detection is rapid, simple, and relatively flexible. Here, 18F-labelled synthesis products were analyzed by TLC and the radioactivity distribution on the plates determined using the three techniques. Radioactivity scanning is appropriate only with good chromatographic resolution and previously validated scanning parameters. Film autoradiography exhibits poor linearity if radioactivity varies greatly. PSL provides high sensitivity and resolution and superior linearity compared with the other methods.
Keywords
TLC , Film autoradiography , Radioactivity scanning , Phosphoimaging technique , Photostimulated luminescence , Planar chromatography , Fluorine-18
Journal title
Applied Radiation and Isotopes
Serial Year
2006
Journal title
Applied Radiation and Isotopes
Record number
548017
Link To Document