Title of article
Standardization of tritiated water by two improved methods
Author/Authors
D. Stânga، نويسنده , , I. Moreau، نويسنده , , P. Cassette، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
4
From page
1203
To page
1206
Abstract
Tritiated water has been standardized in the framework of a French–Romanian cooperation by two improved methods: liquid scintillation counting based on the triple to double coincidence ratio method and the internal gas proportional counting used in conjunction with a tritium generator for chemical reduction of water. The uncertainties of measurement for both methods were smaller than 0.6% and the two results were consistent within these uncertainties, indicating that either method is equally suited for standardizing tritiated water.
Keywords
Triple to double coincidence ratio method , Internal gas counting , Tritiated water standardization