Title of article :
Peak efficiency calibration for attenuation corrected cylindrical sources in gamma ray spectrometry by the use of a point source
Author/Authors :
Julio C. Aguiar، نويسنده , , Eduardo Galiano، نويسنده , , Jorge Fernandez، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
5
From page :
1643
To page :
1647
Abstract :
A theoretical method of determining the gamma-ray peak efficiency for a cylindrical source, based on a modified expression for point sources is derived. A term for the photon self-attenuation is included in the calculation. The method is valid for any source material as long as the source activity concentration is considered to be homogeneous. Results of this expression are checked against experimental data obtained with 241Am, 57Co, 137Cs, and 60Co sources.
Keywords :
Self-attenuation of volume sources , Gamma peak efficiency , HPGe detectors , Efficiency for cylindrical sources
Journal title :
Applied Radiation and Isotopes
Serial Year :
2006
Journal title :
Applied Radiation and Isotopes
Record number :
548121
Link To Document :
بازگشت