Title of article
Validation of analytical formulae for the efficiency calibration of gamma detectors used in laboratory and in-situ measurements
Author/Authors
Mahmoud I. Abbas، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
4
From page
1661
To page
1664
Abstract
Direct analytical formulae for the efficiency calibration of gamma (scintillation and semiconductor) detectors using an arbitrarily positioned radiating point source are extended to include sources with volumetric shapes. The attenuation of photons by the source itself (self-absorption) is determined by calculating the photon path length through the source material. The theoretical and the published experimental efficiency values are in good agreement.
Keywords
Parallelepiped detector , In-situ gamma spectroscopy systems , Extended source and full-energy peak efficiency
Journal title
Applied Radiation and Isotopes
Serial Year
2006
Journal title
Applied Radiation and Isotopes
Record number
548124
Link To Document