• Title of article

    Determination of 222Rn in fresh water: Development of a robust method of analysis by α/β separation liquid scintillation spectrometry

  • Author/Authors

    Jacqueline M. Pates، نويسنده , , Neil J. Mullinger، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    12
  • From page
    92
  • To page
    103
  • Abstract
    Liquid scintillation spectrometry is used widely for determining 222Rn in natural waters; however, the benefits of α/β separation have not been fully explored. The extractants toluene and Ultima Gold F were compared, and both performed well for a range of extreme waters. A robust method for calibrating extraction and counting efficiencies has been developed. Detection limits are 20 mBq l−1 (toluene) and 16 mBq l−1 (UGF) for a 60 min count and 600-ml sample, halving the required sample volume.
  • Keywords
    radon , liquid scintillation , Toluene extraction , Stream–aquifer interaction
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2007
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    548140