• Title of article

    Single-acquisition method for simultaneous determination of extrinsic gamma-camera sensitivity and spatial resolution

  • Author/Authors

    J.A.M. Santos، نويسنده , , S. Sarmento، نويسنده , , P. Alves، نويسنده , , M.C. Torres، نويسنده , , A.L. Bastos، نويسنده , , R. R. Cardoso and C. F. Ponte Neto ، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    6
  • From page
    44
  • To page
    49
  • Abstract
    A new method for measuring simultaneously both the extrinsic sensitivity and spatial resolution of a gamma-camera in a single planar acquisition was implemented. A dual-purpose phantom (SR phantom; sensitivity/resolution) was developed, tested and the results compared with other conventional methods used for separate determination of these two important image quality parameters. The SR phantom yielded reproducible and accurate results, allowing an immediate visual inspection of the spatial resolution as well as the quantitative determination of the contrast for six different spatial frequencies. It also proved to be useful in the estimation of the modulation transfer function (MTF) of the image formation collimator/detector system at six different frequencies and can be used to estimate the spatial resolution as function of the direction relative to the digital matrix of the detector.
  • Keywords
    Gamma-camera , Modulation transfer function , sensitivity , spatial resolution , Linear spread function
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2008
  • Journal title
    Applied Radiation and Isotopes
  • Record number