• Title of article

    X-ray imaging with the PILATUS 100k detector

  • Author/Authors

    M. Bech، نويسنده , , O. Bunk، نويسنده , , C. David، نويسنده , , Rensselaer Polytechnic Institute Russell P. Kraft، نويسنده , , C. Br?nnimann، نويسنده , , E.F. Eikenberry، نويسنده , , F. Pfeiffer، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    5
  • From page
    474
  • To page
    478
  • Abstract
    We report on the application of the PILATUS 100K pixel detector for medical imaging. Experimental results are presented in the form of X-ray radiographs using standard X-ray absorption contrast and a recently developed phase contrast imaging method. The results obtained with the PILATUS detector are compared to results obtained with a conventional X-ray imaging system consisting of an X-ray scintillation screen, lens optics, and a charge coupled device. Finally, the results for both systems are discussed more quantitatively based on an image power spectrum analysis.
  • Keywords
    x-ray imaging , X-ray phase contrast , Pixel detector , direct conversion
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2008
  • Journal title
    Applied Radiation and Isotopes
  • Record number