• Title of article

    Development of a low-level background gamma-ray spectrometer by KRISS

  • Author/Authors

    K.B. Lee، نويسنده , , Tae Soon Park، نويسنده , , Jong Man Lee، نويسنده , , Phil-Je Oh، نويسنده , , Sang-Han Lee، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    5
  • From page
    845
  • To page
    849
  • Abstract
    A new low-level background and high-efficiency gamma-ray spectrometric system, to be used mainly for the activity certification of natural-matrix certified reference materials (CRMs) and environmental reference materials (RMs) that has been developed on the grounds of the Korea Research Institute of Standards and Science (KRISS). The spectrometer consists of a low-background high-purity germanium detector with a relative efficiency of 120% and various shielding devices to reduce radiation background. The cabinet-shaped device made of 10 ton of shielding materials encloses the germanium detector for protection against background from natural radioactivity and neutrons. Three plates of 50-mm-thick plastic scintillation detectors on top of the passive shielding cabinet suppress cosmogenic background by detecting high-energetic cosmic muons bombarding the germanium detector. The measured background rate of the spectrometer for the energy range 50–3000 keV was 1.72 s–1.
  • Keywords
    Low-level background , Anticoincidence shield , Background reduction , Gamma-ray spectrometer
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2008
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    548460