Title of article :
Cascades of pile-up and dead time
Author/Authors :
S. Pommé، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
Count loss through a cascade of pile-up and dead time is studied. Time interval density-distribution functions and throughput factors are presented for counters with a series arrangement of pile-up and extending or non-extending dead time. A counter is considered, where an artificial dead time is imposed on every counted event, in order to control the length and type of dead time. For such a system, it is relatively easy to determine an average count-loss correction factor via a live-time clock gated by the imposed dead-time signal (‘live-time mode’), or otherwise to apply a correction factor based on the inversion of the throughput function (‘real-time mode’). However, these techniques do not account for additional loss through pulse pile-up. In this work, counting errors associated with neglecting cascade effects are calculated for measurements in live-time and real-time mode.
Keywords :
Pulse pile-up , Radionuclide metrology , Counting statistics , Dead time , Nuclear counting
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes