    Luiz Tauhata، نويسنده , , Akira Iwahara، نويسنده , , Antonio E. De Oliveira، نويسنده , , Eduarda Alexandre Rezende، نويسنده , , José Ubiratan Delgado، نويسنده , , Carlos José da Silva، نويسنده , , Joyra A. dos Santos، نويسنده , , Ieda G. N?coli، نويسنده , , Frederico G. Alabarse، نويسنده , , Ana Maria Xavier، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    7
  • From page
    981
  • To page
    987
  • Abstract
    Proficiency tests were applied to assess the performance of 74 nuclear medicine services in activity measurements of 131I, 123I, 99Tcm, 67Ga and 201Tl. These tests produced 913 data sets from comparison programmes promoted by the National Laboratory for Ionizing Radiation Metrology (LNMRI) from 1999 to 2006. The data were evaluated against acceptance criteria for accuracy and precision and assigned as Acceptable or Not acceptable accordingly. In addition, three other statistical parameters were used as complementary information for performance evaluation which related to normative requirements and to radionuclide calibrators. The results have shown a necessity to improve quality control procedures and unsatisfactory performances of radionuclide calibrators, which incorporate Geiger–Müller detectors.
  • Keywords
    Radionuclide calibrators , Intercomparison , Proficiency test , Nuclear medicine services