Title of article :
An approximation for response function to γ-rays of NaI(Tl) detectors up to 1.5 MeV
Author/Authors :
Ahmet Cengiz، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
6
From page :
1371
To page :
1376
Abstract :
The response functions of a 7.62×7.62 cm NaI(Tl) scintillation detector to photons from point γ-ray sources, 10 cm from the scintillator surface, in the energy up to 1.5 MeV, were calculated using the Monte Carlo method, applying simple approximations based on the peak to total ratio and the detector resolution. The Compton continuum of the detector response function was assumed as an isotropic (rectangular) region for the photon energies up to 1 MeV. In the energies between 1 and 1.5 MeV, the Compton continuum was obtained assuming a single Compton scattering with free electrons. The photopeak of the detector response function was assumed as a line. Each determined channel of the response function was distributed to Gaussian functions. The obtained response functions were compared with the experimental values and a good agreement was found.
Keywords :
response function , NaI(Tl) detector , Monte Carlo method , Peak to total ratio , Detector resolution
Journal title :
Applied Radiation and Isotopes
Serial Year :
2008
Journal title :
Applied Radiation and Isotopes
Record number :
548545
Link To Document :
بازگشت