Title of article :
Comparison of doses calculated by the Monte Carlo method and measured by LiF TLD in the buildup region for a 60Co photon beam
Author/Authors :
M. Budanec، نويسنده , , ?. Kne?evi?، نويسنده , , T. Bokuli?، نويسنده , , I. Mr?ela، نويسنده , , M. Vrtar، نويسنده , , B. Veki?، نويسنده , , Z. Kusi?، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
This work studied the percent depth doses of 60Co photon beams in the buildup region of a plastic phantom by LiF TLD measurements and by Monte Carlo calculations. An agreement within ±1.5% was found between PDDs measured by TLD and calculated by the Monte Carlo method with the TLD in a plastic phantom. The dose in the plastic phantom was scored in voxels, with thickness scaled by physical and electron density. PDDs calculated by electron density scaling showed a better match with image; the difference is within ±1.5% in the buildup region for square and rectangular field sizes.
Keywords :
Dose buildup , LiF TLD , Monte Carlo simulation
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes