Title of article :
Verification of dose delivery for a prostate sIMRT treatment using a SLIC–EPID
Author/Authors :
Mohammad-Mohammadi، Alireza نويسنده , , Eva Bezak، نويسنده , , Paul Reich، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
The current work focuses on the verification of transmitted dose maps, measured using a scanning liquid ionization chamber–electronic portal imaging device (SLIC–EPID) for a typical step-and-shoot prostate IMRT treatment using an anthropomorphic phantom at anterior–posterior (A–P), and several non-zero gantry angles. The dose distributions measured using the SLIC–EPID were then compared with those calculated in the modelled EPID for each segment/subfield and also for the corresponding total fields using a gamma function algorithm with a distance to agreement and dose difference criteria of 2.54 mm and 3%, respectively.
Keywords :
dose verification , SLIC–EPID , Segmented IMRT , Transmitted dose
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes