• Title of article

    Hemodynamic and clinical impact of prosthesis–patient mismatch in the aortic valve position and its prevention

  • Author/Authors

    Philippe Pibarot، نويسنده , , Jean G. Dumesnil، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    11
  • From page
    1131
  • To page
    1141
  • Abstract
    Prosthesis–patient mismatch is present when the effective orifice area of the inserted prosthetic valve is less than that of a normal human valve. This is a frequent problem in patients undergoing aortic valve replacement, and its main hemodynamic consequence is the generation of high transvalvular gradients through normally functioning prosthetic valves. The purposes of this report are to present an update on the concept of aortic prosthesis–patient mismatch and to review the present knowledge with regard to its impact on hemodynamic status, functional capacity, morbidity and mortality. Also, we propose a simple approach for the prevention and clinical management of this phenomenon because it can be largely avoided if certain simple factors are taken into consideration before the operation.
  • Keywords
    Left ventricular , BSA , body surface area , EOA , effective orifice area , LV
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Serial Year
    2000
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Record number