Title of article
Hemodynamic and clinical impact of prosthesis–patient mismatch in the aortic valve position and its prevention
Author/Authors
Philippe Pibarot، نويسنده , , Jean G. Dumesnil، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
11
From page
1131
To page
1141
Abstract
Prosthesis–patient mismatch is present when the effective orifice area of the inserted prosthetic valve is less than that of a normal human valve. This is a frequent problem in patients undergoing aortic valve replacement, and its main hemodynamic consequence is the generation of high transvalvular gradients through normally functioning prosthetic valves. The purposes of this report are to present an update on the concept of aortic prosthesis–patient mismatch and to review the present knowledge with regard to its impact on hemodynamic status, functional capacity, morbidity and mortality. Also, we propose a simple approach for the prevention and clinical management of this phenomenon because it can be largely avoided if certain simple factors are taken into consideration before the operation.
Keywords
Left ventricular , BSA , body surface area , EOA , effective orifice area , LV
Journal title
JACC (Journal of the American College of Cardiology)
Serial Year
2000
Journal title
JACC (Journal of the American College of Cardiology)
Record number
596107