Title of article :
Pressure rinsing of coronary stents immediately before implantation reduces inflammation and neointimal hyperplasia
Author/Authors :
Antoni Bayes-Genis، نويسنده , , Allan R. Camrud RN، نويسنده , , Michael Jorgenson، نويسنده , , Janis Donovan، نويسنده , , Kristin L. Shogren، نويسنده , , David R. Holmes Jr، نويسنده , , Robert S. Schwartz، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
7
From page :
562
To page :
568
Abstract :
OBJECTIVES This study evaluates whether rinsing stents with high pressure immediately before implantation minimizes stent-induced inflammation and neointimal formation. BACKGROUND Several reports indicate that manual stent manipulation before implantation results in foreign body contamination and increased neointimal hyperplasia. METHODS A stent-cleaning chamber was developed to rinse stents at a sustained hydrodynamic pressure of 4 atm for 10 s. Commercial pre-mounted stents were examined with different levels of manipulation: 1) untouched stents: no stent manipulation before implantation; 2) handled stents: manual stent re-crimping on the balloon; 3) rinsed stents: pressure-rinsed with the stent-cleaning chamber. In vitro surface analysis was evaluated by scanning electron microscopy. Neointimal hyperplasia and inflammation around stent struts were also assessed in the pig in-stent restenosis model. RESULTS In vitro analysis revealed fewer contaminants on rinsed stents compared with untouched (p = 0.01) and handled stents (p < 0.001). In vivo, neointimal thickness, neointimal area and vessel percent stenosis were significantly reduced in rinsed, compared with not-rinsed, stents (p = 0.002, P = 0.007, P = 0.008 respectively). In addition, a significant reduction in the inflammatory infiltrate around struts was observed in untouched, compared with handled, stents (p = 0.04) and in rinsed, compared with not-rinsed, stents (p < 0.001). Regression analysis accounting for injury and neointimal thickness showed significant differences in slopes between “handled + not-rinsed” and “handled + rinsed” stents (p = 0.004), and between “untouched + not-rinsed” and “untouched + rinsed stents” (p = 0.037). CONCLUSIONS Rinsing stents under high pressure immediately before coronary implantation results in less inflammation around struts and thinner neointima at 28 days in this pig model.
Keywords :
ANOVA , Analysis of variance , IM , Intramuscular , MMA , Methylmethacrylate
Journal title :
JACC (Journal of the American College of Cardiology)
Serial Year :
2001
Journal title :
JACC (Journal of the American College of Cardiology)
Record number :
596749
Link To Document :
بازگشت