Title of article :
A prospective, double-blind study on the efficacy of the bioline surface-heparinized extracorporeal perfusion circuit
Author/Authors :
George M. Palatianos، نويسنده , , Christophoros N. Foroulis، نويسنده , , Maria I. Vassili، نويسنده , , George Astras، نويسنده , , Konstantinos Triantafillou، نويسنده , , Emmanuel Papadakis، نويسنده , , Angela A. Lidoriki، نويسنده , , Eugenia Iliopoulou، نويسنده , , Efthimia N. Melissari، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
7
From page :
129
To page :
135
Abstract :
Background We evaluated the newly introduced Bioline heparin coating and tested the hypothesis that surface heparinization limited to the oxygenator and the arterial filter will ameliorate systemic inflammation and preserve platelets during cardiopulmonary bypass (CPB). Methods In a prospective double-blind study, 159 patients underwent coronary revascularization using closed-system CPB with systemic heparinization, mild hypothermia (33°C), a hollow-fiber oxygenator, and an arterial filter. The patients were randomly divided in three groups. In group A (controls, N = 51), surface heparinization was not used. In group B (n = 52), the extracorporeal circuits were totally surface-heparinized with Bioline coating. In group C (n = 56), surface heparinization was limited to oxygenator and arterial filter. Results No significant difference was noted in patient characteristics and operative data between groups. Operative (30-day) mortality was zero. Platelet counts dropped by 12.3% of pre-CPB value among controls at 15 minutes of CPB, but were preserved in groups B and C throughout perfusion (p = 0.0127). Platelet factor 4, plasmin-antiplasmin levels, and tumor necrosis factor-α increased more in controls during CPB than in groups B or C (p = 0.0443, p = 0.0238 and p = 0.0154 respectively). Beta-thromboglobulin, fibrinopeptide-A, prothrombin fragments 1 + 2, factor XIIa levels, bleeding times, blood loss, and transfusion requirements were similar between groups. Intensive care unit stay was shorter in groups B and C than in controls (p = 0.037). Conclusions Surface heparinization with Bioline coating preserves platelets, ameliorates the inflammatory response and is associated with a reduced fibrinolytic activity during CPB. Surface heparinization limited to the oxygenator and the arterial filter had similar results as totally surface-heparinized circuits.
Journal title :
The Annals of Thoracic Surgery
Serial Year :
2003
Journal title :
The Annals of Thoracic Surgery
Record number :
606736
Link To Document :
بازگشت