Kiyotaka Fukamachi، نويسنده , , Masahiro Inoue، نويسنده , , Zoran B. Popovi ، نويسنده , , Kazuyoshi Doi، نويسنده , , Soren Schenk، نويسنده , , Hassan Nemeh، نويسنده , , Yoshio Ootaki، نويسنده , , Michael W. Kopcak Jr، نويسنده , , Raymond Dessoffy، نويسنده , , James D. Thomas، نويسنده , , Richard W. Bianco، نويسنده , , James M. Berry، نويسنده , , Patrick M. McCarthy، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
5
From page
688
To page
692
Abstract
Purpose
The purpose of this study was to evaluate the ability of the Myocor Coapsys device to restore leaflet apposition and valve competency off-pump in a canine model of functional mitral regurgitation (MR).
Description
The Coapsys device was surgically implanted in 10 dogs after MR induction by rapid ventricular pacing. The Coapsys consists of anterior and posterior epicardial pads connected by a subvalvular chord. The annular head of the posterior pad was positioned at the annular level to draw the posterior leaflet and annulus toward the anterior leaflet. Final device size was selected when MR was minimized or eliminated as assessed by color flow Doppler echocardiography.
Evaluation
All implants were placed off-pump without atriotomy, and mean MR grade was reduced from 2.9 ± 0.7 to 0.6 ± 0.7 (p< 0.001) acutely. No hemodynamic compromise was noted.
Conclusions
The Coapsys device consistently and significantly reduced or eliminated functional MR acutely. Further study will be required to assess the chronic stability of the repair in this animal model.