• Title of article

    Monitoring weaning from BIVAD thoratec with peak oxygen consumption

  • Author/Authors

    Pascal H. Colson، نويسنده , , Frederique Ryckwaert، نويسنده , , Max Saussine، نويسنده , , Marc Ferrière، نويسنده , , Bernard Albat، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    3
  • From page
    1808
  • To page
    1810
  • Abstract
    A biventricular assistance device has been implanted in a young woman for a peripartum cardiac failure. An intended weaning consisted of gradual reloading and exercise training monitored with peak oxygen consumption (V 2) and radionuclide-left ventricle ejection fraction. Progressive increase in peak V 2during partial assistance occurred more than 2 months, from 10.3 to 19 mL • kg−1 • min−1. Successful explantation was realized when peak V 2 exceeded 15 mL • kg−1 • min−1 and radionuclide-left ventricle ejection fraction was more than 40% during off-pump testing.
  • Journal title
    The Annals of Thoracic Surgery
  • Serial Year
    2004
  • Journal title
    The Annals of Thoracic Surgery
  • Record number

    607568