Title of article :
Aortic Leaflet Replacement With the New 3F Stentless Aortic Bioprosthesis
Author/Authors :
Mirko Doss، نويسنده , , Sven Martens، نويسنده , , Jeffrey P. Wood، نويسنده , , Aleksandra Miskovic، نويسنده , , Triantafyllia Christodoulou، نويسنده , , Gerhard Wimmer-Greinecker، نويسنده , , Anton Moritz، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
4
From page :
682
To page :
685
Abstract :
Purpose Clinical trials with the new 3F stentless aortic bioprosthesis began October 2001, and as one of the first centers to implant this prosthesis in humans, we would like to present our experiences with this new device. Description The 3F aortic bioprosthesis is a stentless biological heart valve fabricated from three equal leaflets of equine pericardium, assembled in a tubular shape, and implanted in the native aortic root to replace the patientʹs diseased aortic leaflets. Between January 2002 and August 2002, 24 3F aortic bioprostheses were implanted at our institution. Effective orifice area, mean gradients, and ejection fraction were evaluated by echocardiography at discharge and at 12-month follow-ups after surgery. Evaluation At 12-month follow-ups, the 3F bioprosthesis showed a good hemodynamic performance with a significant drop of mean gradients to 10.3 mm Hg, a mean effective orifice area of 1.7 cm2, and a mean ejection fraction of 61.5%. Conclusions The clinical performance of the new 3F aortic bioprosthesis is comparable with regular stentless aortic valves. However its unique design facilitates implantation.
Journal title :
The Annals of Thoracic Surgery
Serial Year :
2005
Journal title :
The Annals of Thoracic Surgery
Record number :
608345
Link To Document :
بازگشت