Title of article :
The Second-Generation Aortic Connector: Six Months’ Angiographic Follow-Up
Author/Authors :
Marek Setina، نويسنده , , Adriana Krchnakova، نويسنده , , Ales Mokracek، نويسنده , , Ladislav Pesl، نويسنده , , Luis F. Bonilla، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
5
From page :
695
To page :
699
Abstract :
Purpose Recently, new mechanical anastomotic devices have been developed. Initial results appear to be equivalent to those obtained with suture. The aim of the study was to evaluate the 6-month angiographic patency and clinical results with the St. Jude Medical second-generation aortic connector for proximal aortosaphenous graft anastomosis. Description From September 2002 to June 2003, 45 connectors were implanted in 39 patients. Thirty-three patients with 36 connectors underwent 6-month angiographic and clinical follow-up. Evaluation One connector had an early occlusion and 2 connectors and 1 vein graft were occluded at 6-month angiography, for a patency rate of 88.9% (32 of 36). No device-related complications were detected at 6-month follow-up. Conclusions The second-generation aortic connector is safe and easy to use. Preliminary results show no device-related complications and a satisfactory 6-month angiographic patency.
Journal title :
The Annals of Thoracic Surgery
Serial Year :
2005
Journal title :
The Annals of Thoracic Surgery
Record number :
608885
Link To Document :
بازگشت