Title of article
Reinforcement of Left Ventricular Assist Device Outflow Grafts to Prevent Kinking
Author/Authors
William E. Cohn، نويسنده , , Igor D. Gregoric، نويسنده , , OH Frazier، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
2
From page
301
To page
302
Abstract
Because of their small size, rotary left ventricular assist devices (LVADs) can be implanted through several alternative approaches, including transdiaphragmatic LVAD insertion through a left subcostal incision with anastomosis of the outflow graft to the retroperitoneal supra-celiac aorta and a left thoracotomy approach with anastomosis to the descending thoracic aorta. More recently we have added a counter-incision to allow the outflow graft of a transdiaphragmatic LVAD to be tunneled through the right chest and anastomosed to the ascending aorta. However, constructing a tension-free, nonkinking lie of the outflow graft can be challenging. We have found that placing a 10-cm to 15-cm length of polytetrafluoroethylene graft coaxially over the LVAD outflow graft ensures a smooth, kink-free lie. Thus far 12 patients have undergone reinforcement of the LVAD outflow graft prior to graft-to-aorta anastomosis. In all cases, graft lie was facilitated and kinking was eliminated.
Journal title
The Annals of Thoracic Surgery
Serial Year
2007
Journal title
The Annals of Thoracic Surgery
Record number
610840