Title of article :
Seven-Day Artificial Lung Testing in an In-Parallel Configuration
Author/Authors :
Hitoshi Sato، نويسنده , , Grant W. Griffith، نويسنده , , Candice M. Hall، نويسنده , , John M. Toomasian، نويسنده , , Ronald B. Hirschl، نويسنده , , Robert H. Bartlett، نويسنده , , Keith E. Cook، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
Background
A thoracic artificial lung, the MC3 Biolung, is being developed as a bridge to lung transplantation or as a treatment for acute respiratory insufficiency.
Methods
The thoracic artificial lung was tested in 10 sheep with the goal of 7 days of respiratory support. The sheep were recovered from surgery and monitored awake for 7 days. Hemodynamics, blood gases, blood cell counts, and organ function were recorded, and after 7 days, all sheep were euthanized and necropsied.
Results
Seven sheep survived the full duration. Cardiac output and mean arterial blood pressure were unchanged, averaging 4.7 ± 0.8 L/min and 98 ± 10 mm Hg, respectively. Arterial oxygen tension and device oxygen transfer rate were also unchanged, averaging 110 ± 26 mm Hg and 97.7 ± 35 mL/min, respectively. Arterial carbon dioxide tension was within normal ranges during the entire experiment, averaging 37.4 ± 3.8 mm Hg. Artificial lung blood flow decreased from 51% ± 14% of cardiac output on day 1 to 30% ± 16% by day 7 because of changes in natural and artificial lung resistance. White blood cell counts were significantly elevated on days 5 and 7, and lastly, kidney and liver function remained normal, although signs of kidney infarction or hemorrhage were noted.
Conclusions
The thoracic artificial lung is suitable for 7-day attachment, but improvements in blood biocompatibility are warranted.