Title of article :
Corrosion of Pure Titanium Sternal Wire
Author/Authors :
Yasuko Tomizawa، نويسنده , , Takao Hanawa، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
3
From page :
1012
To page :
1014
Abstract :
Titanium and its alloys are considered reliable materials commonly used in heart surgeries. A patient had severe inflammation at the median sternotomy wound develop after open heart surgery. Sternal wires were removed and were found to be seriously corroded on most of the surface. Energy dispersive spectroscopy identified titanium as the only elemental component of the explanted wire with grade 2 purity, and the polarization curve showed satisfactory corrosion resistance similar to commercial pure titanium. Prolonged inflammation might have caused extensive corrosion of titanium in the present case.