• Title of article

    A Novel Adjustable Pulmonary Artery Banding System for Hypoplastic Left Heart Syndrome

  • Author/Authors

    Renato S. Assad، نويسنده , , Marina M. Zamith، نويسنده , , Maria Fernanda Silva، نويسنده , , Petrônio G. Thomaz، نويسنده , , Leonardo A. Miana، نويسنده , , Vitor C. Guerra، نويسنده , , Carlos A. Pedra، نويسنده , , Miguel B. Marcial، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    4
  • From page
    2081
  • To page
    2084
  • Abstract
    Purpose We describe the first case in which a neonate with hypoplastic left heart syndrome was initially managed using a mini adjustable banding system. Description Through a mid-sternotomy, a 5-day-old neonate underwent bilateral pulmonary artery banding using this new system, combined with placement of a main pulmonary artery to the innominate artery shunt. Evaluation The patient had an uneventful postoperative course. Three percutaneous adjustments of the banding system were necessary to keep the arterial oxygen saturation in the 75% to 85% range. On day 48 of life, she was submitted to stent placement (6 mm) within the atrial septum to treat a restrictive atrial septal defect. Afterward, seven additional percutaneous adjustments of the banding system were necessary. The Norwood operation and the bidirectional Glenn shunt were carried out on the day 106 of life. The bands were removed with no pulmonary artery distortion. Conclusions The clinical use of this innovative pulmonary artery banding system was feasible, safe, and effective. This allowed for customization of the pulmonary blood flow according to the underlying clinical needs, resulting in a more precise balance between the pulmonary and systemic circulations.
  • Journal title
    The Annals of Thoracic Surgery
  • Serial Year
    2007
  • Journal title
    The Annals of Thoracic Surgery
  • Record number

    611208