• Title of article

    Traumatic Dehiscence of Medtronic Freestyle Stentless Bioprosthesis

  • Author/Authors

    Turki Albacker، نويسنده , , Adil Al Kindi، نويسنده , , Benoit de Varennes، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    3
  • From page
    643
  • To page
    645
  • Abstract
    We report a case of deceleration injury causing traumatic dehiscence of a Medtronic Freestyle aortic bioprosthesis (Medtronic Inc, Minneapolis, MN) in a 70-year-old patient 5 years after a total root reconstruction. The patient presented with progressive exertional dyspnea 6 months after a significant impact during a hockey game. Intraoperatively, there was a posterior dehiscence of the proximal suture line and almost complete tear of the left coronary leaflet. At reoperation, the patient underwent reconstruction of the aortic root and Bentall procedure with a size 23 mechanical aortic valved conduit.
  • Journal title
    The Annals of Thoracic Surgery
  • Serial Year
    2008
  • Journal title
    The Annals of Thoracic Surgery
  • Record number

    611834