• Title of article

    Low thromboembolic risk without anticoagulation using advanced-design left ventricular assist devices

  • Author/Authors

    James P. Slater، نويسنده , , Eric A. Rose، نويسنده , , Howard R. Levin، نويسنده , , OH Frazier، نويسنده , , J. Kirk Roberts، نويسنده , , Alan D. Weinberg، نويسنده , , Mehmet C. Oz، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    8
  • From page
    1321
  • To page
    1328
  • Abstract
    Background. A major limitation of cardiac assist devices has been the high incidence of thromboembolic events and their requirement for systemic anticoagulation. The Thermo Cardiosystems HeartMate 1000 IP left ventricular assist device (LVAD) employs a design that may reduce thromboembolic risk and obviate the need for systemic anticoagulation. Methods. Two hundred twenty-three patients with nonreversible heart failure were supported with the HeartMate LVAD as a bridge to heart transplantation. All patients were monitored prospectively for thromboembolic events. Anticoagulation regimens and occurrence of subclinical thromboembolic events, including those seen by transcranial Doppler examinations in selected patients, were also recorded. Results. Total time of LVAD support use was 531.2 patient-months. Twenty-three patients (10%) received warfarin postoperatively for 42.4 patient-months (8.2% of total support time). Six patients (2.7%) had thromboembolic events, representing 0.011 events per patient-month of device use. Conclusions. The thromboembolic complication rate associated with this LVAD is acceptably low despite the minimal anticoagulation employed in this series, allowing consideration of long-term device use for the treatment of heart failure.
  • Journal title
    The Annals of Thoracic Surgery
  • Serial Year
    1996
  • Journal title
    The Annals of Thoracic Surgery
  • Record number

    613814