Title of article :
Single-Incision and Single-Element Array Electrode to Lower the Defibrillation Threshold
Author/Authors :
Volker Kühlkamp MD، نويسنده , , Kourosh Khalighi MD، نويسنده , , Volker D?rnberger MD، نويسنده , , Gerhard Ziemer MD، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Abstract :
Occasional patients have excessive defibrillation energy requirements despite appropriate transvenous defibrillation lead position and the use of biphasic shocks. A single-element subcutaneous array electrode was implanted in 2 patients with a high defibrillation threshold. The array electrode was implanted through the same infraclavicular incision that was used for implantation of the transvenous lead. The defibrillation threshold decreased from 30 J to 15 J and from 24 J to 9 J with the subcutaneous array electrode.
Journal title :
The Annals of Thoracic Surgery
Journal title :
The Annals of Thoracic Surgery