Title of article
Frequency doubling technology perimetry for detection of glaucomatou viual field lo
Author/Authors
Kimberly E. Cello، نويسنده , , Jacqueline M. Nelon-Quigg، نويسنده , , Chri A. Johnon، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
9
From page
314
To page
322
Abstract
PURPOE:
To evaluate the ability of frequency doubling technology perimetry to detect early, moderate, and advanced glaucomatou viual field lo.
METHOD:
In a propective tudy, frequency doubling technology perimetry (C-20 full threhold) wa performed in the right eye of 254 normal control ubject and 230 patient with early (n = 85), moderate (n = 114), or advanced (n = 31) glaucomatou viual field lo. Previou Humphrey Field Analyzer tet reult were ued to claify glaucomatou viual field lo a early (mean deviation no wore than −6 dB), moderate (mean deviation between −6 and −12 dB) or advanced (mean deviation between −12 and −22 dB).
REULT:
Receiver operating characteritic curve howed 100% enitivity and pecificity (area under the curve, 1.0) for detecting advanced glaucomatou viual field lo, approximately 96% enitivity and 96% pecificity (area under the curve, 0.9751) for detecting moderate glaucomatou viual field lo, and approximately 85% enitivity and 90% pecificity (area under the curve, 0.9261) for early glaucomatou viual field lo.
CONCLUION:
Frequency doubling technology perimetry demontrate high enitivity and pecificity for detection of early, moderate, and advanced glaucomatou viual field lo.
Journal title
American Journal of Ophthalmology
Serial Year
2000
Journal title
American Journal of Ophthalmology
Record number
622712
Link To Document