Title of article :
mall-ize pediatric vitrectomy wide-angle contact len
Author/Authors :
Gholam A. Peyman، نويسنده , , Chritina Canaki، نويسنده , , Charalampo Livir-Rallato، نويسنده , , Paul Whalen، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
2
From page :
236
To page :
237
Abstract :
Purpoe We decribe a new wide-field vitrectomy contact len pecifically deigned to facilitate pediatric vitreoretinal urgery. Deign Experimental tudy, intrument decription. Method A wide-field contact two-piece gla vitrectomy len wa deigned with the larget diameter of 13.6 mm and a contact diameter of 7.00 mm, acrificing only the more tilt-enitive part of the optic in order to reduce it phyical ize. Reult Laboratory meaurement in a model eye demontrated a 61-degree binocular field of view, a 94-degree tatic field of view, and a maximum 129-degree field of view. Image magnification wa 0.50×. No complication were linked to the ue of thi intrument, and the need to exchange for another len wa never encountered in 10 cae. The two-piece gla contruction enabled thi len to be terilized in an autoclave or terilized by any other intrument terilization technique. Concluion The intrument uccefully fulfilled the purpoe of it deign, providing at the ame time an excellent wide field of view and ample pace for the urgeon, performing well in 10 pediatric vitreoretinal cae (8 cae of retinopathy of prematurity; 2 cae of traumatic eye injury). It looe two-piece deign rendered it uitable for all terilization technique.
Journal title :
American Journal of Ophthalmology
Serial Year :
2003
Journal title :
American Journal of Ophthalmology
Record number :
624100
Link To Document :
بازگشت