Title of article :
A New Non-trocar ytem for 25-gauge Tranconjunctival Par Plana Vitrectomy
Author/Authors :
Mamoru Tei، نويسنده , , Takaumi himamoto، نويسنده , , Toru Yauhara، نويسنده , , Hideki Komori، نويسنده , , Hideo Oda، نويسنده , , higeru Kinohita، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
Purpoe
To introduce a new non-trocar ytem for 25-gauge tranconjunctival par plana vitrectomy (PPV).
Deign
tudy of a new urgical intrument.
Method
Thi new non-trocar ytem for 25-gauge tranconjunctival PPV conit of a contact len ring featuring four projection containing 1.0-mm diameter hole located 3.0 mm from the ring edge, a wedge-haped 25-gauge infuion cannula, and two plug.
Reult
The ring i fixed with 7-0 ilk uture at the 3- and 9-o’clock poition on the limbu. Uing the 25-gauge needle, three conjunctival and cleral inciion (n = 3) are made at the projection hole located inferotemporally, uperonaally, and uperotemporally. No intra- or potoperative complication were encountered in 14 patient treated by thi non-trocar 25-gauge tranconjunctival PPV.
Concluion
Uing the ytem introduced here, 25-gauge tranconjunctival PPV can be performed without uturing clerotomie and without intra- or potoperative complication.
Journal title :
American Journal of Ophthalmology
Journal title :
American Journal of Ophthalmology