Title of article :
Optical pH Sensor Based on Calcon Immobilization Membrane
Author/Authors :
S. Rastegarzadeh and Z. Moradpour، نويسنده ,
Issue Information :
فصلنامه با شماره پیاپی سال 2009
Pages :
6
From page :
857
To page :
862
Abstract :
An optical pH sensor was developed by immobilizing Calcon on a porous cellulosic polymer film. The color of the membrane in acidic to basic medium changes from pink to blue, which can be used for determination of pH by spectrophotometry. The sensor response to the pH changes at two wavelengths 510 and 670 nm was investigated. The optical sensor can be used for repetitive and reversible pH measurement in the pH range of 4-9 with a response time of 5 min at 510 nm. The relative standard deviation (R.S.D) was less than 0.51% for seven times alternative measurements of pH from 7 to 8. The sensor was successfully applied to the determination of pH in tap and waste water samples
Keywords :
pH optical sensor , Calcon , Methyltrioctylammonium chloride , Triacetylcellulose
Journal title :
Journal of the Iranian Chemical Society (JICS)
Serial Year :
2009
Journal title :
Journal of the Iranian Chemical Society (JICS)
Record number :
666896
Link To Document :
بازگشت