• Title of article

    Ion-Selective Carbon Paste Electrode Based on 2-Amino-N-benzthioazolyl Benzamide (ABTB) for Determination of Copper (II) by Potentiometric Method

  • Author/Authors

    Mahdavinia، G. H نويسنده Islamic Azad University, Marvdasht Branch, Department of Chemistry, Marvdasht, Iran. , , Chaloosi، M نويسنده Faculty of Chemistry, Tarbiat Moallem University, PO Box 15614, Tehran, Iran , , Shirzadmehr، A نويسنده Faculty of Chemistry, Tarbiat Moallem University, PO Box 15614, Tehran, Iran ,

  • Issue Information
    فصلنامه با شماره پیاپی 2 سال 2010
  • Pages
    8
  • From page
    89
  • To page
    96
  • Abstract
    2-Amino-N-benzthioazolyl benzamide (ABTB) was synthesized, characterized and used for the fabrication of a potentiometric sensor for Cu2+ metal ions. The electrode exhibits linear response to Cu (II) over a wide concentration range (4.79×10 -8 – 1.85×10 -1 M) with Nernstian slope of 30 ± 1.5 mV per decade. The electrode can be used in the pH range from 2 to 9. It has a fast response time of about 10 s and can be used for a period of four months with good reproducibility. The detection limit of this electrode was 2.91×10 - 8 M. The proposed electrode shows a very good selectivity for Cu (II) over a wide variety of metal ions. This chemically modified carbon paste electrode was successfully used for the determination of Cu (II) in electronics waste sample solution
  • Journal title
    Journal of Physical Chemistry and Electrochemistry(JPCE)
  • Serial Year
    2010
  • Journal title
    Journal of Physical Chemistry and Electrochemistry(JPCE)
  • Record number

    683458