• Title of article

    Coated Wire Lead Ion Selective Electrode Based on Dithizone

  • Author/Authors

    Elsalamouny، Ahmed Ragaa نويسنده Faculty of Science, Al-azhar University, Cairo, Egypt , , Elreefy، Sohair Abd-Elfatah نويسنده Hot Labs Center, Atomic Energy Authority, Cairo, Egypt , , Aly Hassan، Aly Mustafa نويسنده Faculty of Science, Al-azhar University, Cairo, Egypt ,

  • Issue Information
    دوماهنامه با شماره پیاپی سال 2012
  • Pages
    8
  • From page
    178
  • To page
    185
  • Abstract
    PVC based membrane of 1,5-diphenylthiocarbazone (dithizone) reveals a Nernstain potentiometric response with the slope of 29+2 mV per decade for Pb2+ over a concentration range (5.0×10-6-1.0×10-2 M). The used electrode was suitable in aqueous solutions in a pH range of 8 to 10. The response time of the electrode is about 15 s and was used for a period of 45 days. The proposed electrode was used in two different applications such as the determination of lead concentration in a part of lead water pipe and as an indicator electrode for potentiometric titration of Pb2+ ions.
  • Journal title
    Analytical and Bioanalytical Electrochemistry
  • Serial Year
    2012
  • Journal title
    Analytical and Bioanalytical Electrochemistry
  • Record number

    691036