Title of article :
High Frequency Electromagnetic Interference Reduction on the SQUID Attached in Pulse Tube Cryocooler
Author/Authors :
Eshraghi، M.J نويسنده , , Cho، M نويسنده , , Kim، JM نويسنده , , Sasada، I نويسنده ,
Issue Information :
فصلنامه با شماره پیاپی 0 سال 2008
Pages :
4
From page :
33
To page :
36
Abstract :
Measurements of the high frequency electromagnetic interference (EMI) of a pulse tube cryocooler are made in a typical laboratory environment using laboratory-made current transformers. The inverter makes high frequency EMI which could cause severe interference with SQUID operation inside the pulse tube cryocooler. By covering the power lines, including a neutral line, of the valve motor with a copper mesh cover, as a ground line, one order de- crease in the magnitude of the EMI noise is successfully obtained. This paper describes an example to show how to observe and reduce conducted high frequency noise currents in a precision experimental system.
Journal title :
Journal of Theoretical and Applied Physics
Serial Year :
2008
Journal title :
Journal of Theoretical and Applied Physics
Record number :
691223
Link To Document :
بازگشت