Title of article :
Performance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using GATE
Author/Authors :
گرامي فر، پرهام نويسنده Faculty of Nuclear Engineering and Physics, Amir Kabir University of Technology Geramifar, Parham , آي، محمدرضا نويسنده دانشگاه علوم پزشكي تهران: مركز تحقيقات علوم و تكنولوژي در پزشكي تهران Ay, Mohammad Rreza , شمسايي ظفرقندي، مجتبي نويسنده دانشگاه صنعتي اميركبير,دانشكده مهندسي هسته اي و فيزيك , , لدوس، جرج نويسنده 5Department of Medical Instruments Technology, Technological Educational Institute, Athens, Greece Loudos, George , رحميم، آرمان نويسنده دانشگاه علوم پزشكي جان هاپكينز بالتي مور ,
Issue Information :
فصلنامه با شماره پیاپی 32 سال 2009
Abstract :
Combined PET/CT scanners now play a major role in medicine for in vivo imaging in oncology, cardiology,
neurology, and psychiatry. As the performance of a scanner depends not only on the scintillating material but also
on the scanner design, with regards to the advent of newer scanners, there is a need to optimize acquisition protocols
as well as to compare scanner performances on an objective basis. In this study we evaluate and compare the
performance of 4 Commercial GE PET/CT cameras, the (i) BGO-based Discovery LS PET/CT (DLS), (ii) the BGObased
Discovery ST PET/CT (DST), (iii) the BGO-based Discovery STE PET/CT (DSTE) and finally (iv) the
LYSO-based Discovery RX PET/CT (DRX) scanner using the Geant4 Application for Tomographic Emission
(GATE). GATE is an open source Monte Carlo simulation platform developed for PET and SPECT studies and is
supported by the OpenGATE collaboration. In accordance with the National Electrical Manufactures Association
(NEMA) NU 2-2001 protocols, the validation of models is carried out against actual published measurements and
the performance comparison is done for sensitivity, scatter fraction and count rate performance, showing very
similar performance compared with published results, thus enabling investigations to better model system
performance (e.g. resolution degradation) within the reconstruction task.. The simulated results demonstrate highest
sensitivity performance with the DST (though with the highest scatter fraction), and highest NECR performance for
the LYSO-based DRX, The results also show that DRX, DLS and DSTE PET/CT cameras have nearly the same
amount of scatter fraction.