Abstract :
We have developed a simple and portable technique for measuring moderately high levels of 222Rn (t1/2 = 3.8 d) in natural waters such as coastal water, groundwater, and river water. The water sample is carefully collected in a glass bottle, and the sample bottle is connected to a radon-in-air monitor in a closed air-loop mode. By purging air through the sample, radon is emanated from the water until a chemical equilibration is obtained between the two phases. The radon in the air loop is determined using the radon-in-air monitor. Then, the radon in water is calculated by a radon-partitioning factor between water and air for a measured water temperature. This technique is especially convenient for determination of 222Rn in natural waters on field sites, since it eliminates the preparation of He gas, cold traps, and alpha-scintillation cells and counter, which are required for traditional radon emanation methods.
Keywords :
natural waters , Radon monitor , Radon measurement , groundwater , Rn exchanger , coastal water , Rn-222