Title of article
An interlaboratory round robin study on the analysis of toxaphene in a cod liver oil standard reference material
Author/Authors
P. Andrews، نويسنده , , K. Headrick، نويسنده , , J-C. Pilon، نويسنده , , B. Lau، نويسنده , , D. Weber، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
10
From page
4393
To page
4402
Abstract
An inter-laboratory round robin analysis of toxaphene in a National Institute of Standards Cod Liver Oil #1588 was carried out. Analysis was performed by GC-ECD, electron capture negative ion GCMS (ECNI GCMS) and electron impact GC/MS in high resolution single ion monitoring mode (EI-HRSIM). Results of the analyses by GC-ECD, ECNI and HRSIM were 4.22±2.45 ppm, 3.95±1.57 ppm, and 2.35±0.06 ppm respectively. Due to the wide inter-laboratory variation in workup, GC conditions, detection method, and quantitation algorithms used, no one set of factors stood out as the cause of the variation in results.
Journal title
Chemosphere
Serial Year
1995
Journal title
Chemosphere
Record number
722541