• Title of article

    Technique for aerosol generation with controllable micrometer size distribution Original Research Article

  • Author/Authors

    Yao-Chuan Lee، نويسنده , , Fu-Tien Jeng، نويسنده , , Chih-Chieh Chen، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    8
  • From page
    760
  • To page
    767
  • Abstract
    The purpose of this study is to develop an aerosol generating system that can produce particles of micrometer size in a convenient and efficient way. This system is comprised of an ultrasonic atomizer, potassium sodium tartrate tetrahydrate (PST) as solute and a program-controlled solute feeding unit with different PST concentrations. Both the aerosol concentration and size distribution pattern can be easily controlled and reproduced in the developed system. While the initial size of droplets generated from atomizer may remain unchanged, the size of residual dry aerosols was controlled by the solute concentration adjusted by the mixing ratio of solute and water. In addition, PST concentration could be alternatively adjusted in any cyclic way to provide particles with relatively mono-disperse, bimodal, varying size as well as skew distribution to meet requirements for various applications. The main advantage of the generating system is to generate particles of specific size distribution in order to simulate aerosols in ambient air or working places.
  • Keywords
    Aerosol generatorUltrasonic atomizerSize distribution patternTest aerosol
  • Journal title
    Chemosphere
  • Serial Year
    2008
  • Journal title
    Chemosphere
  • Record number

    726531