Title of article :
Characterization of a particle trap impactor
Author/Authors :
D.S. Kim، نويسنده , , K.W. Lee، نويسنده , , Y.J. Kim، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2006
Abstract :
A particle trap impactor with a cup shaped collector was evaluated with varying geometrical and operating parameters. The acceleration nozzles having different diameters were designed, and the distance between the acceleration nozzle and the particle trap was changed. The aim of this work was to find out the accurate design parameters of particle trap impactors. Interestingly, our results indicated that (i.e., of a particle having a 50% probability of separation) of particle trap impactors were about 0.6, whereas those of inertial impactors and virtual impactors were 0.49 and 0.69, respectively. It is apparent that values of particle trap impactors can be set at 0.6, even though their design conditions are somewhat different. In addition, the penetration efficiency was found to be dependent on the ratio of nozzle-to-trap distance to nozzle diameter, the sampling flow rate and the nozzle diameter.
Keywords :
Particle trap impactor , Nozzle-to-trap distance , Penetration , Nozzle diameter
Journal title :
Journal of Aerosol Science
Journal title :
Journal of Aerosol Science