Title of article
Time-resolved Hadamard fluorescence imaging
Author/Authors
Hassler، Kai نويسنده , , Anhut، Tiemo نويسنده , , Lasser، Theo نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
-7563
From page
7564
To page
0
Abstract
We present a new concept for fluorescence lifetime imaging (FLIM) based on time-resolved Hadamard imaging (HI). HI allows image acquisition by use of one single-point detector without requiring a moving scanning stage. Moreover, it reduces the influence of detector noise compared with raster scanning. By use of Monte Carlo simulations it could be confirmed that Hadamard transformation may decrease the error in lifetime estimation and in general in fluorescence parameter estimation when the signal-to-noise ratio is low and detector dark noise is high. This concept may find applications whenever the performance of FLIM or similar methods is limited by high dark-count rates and when the use of a single-point detector is preferable.
Keywords
Transforms , Medical optics , Biotechnology , Fluorescence microscopy , Imaging systems
Journal title
Applied Optics
Serial Year
2005
Journal title
Applied Optics
Record number
75075