Abstract :
A study was conducted for the purpose of improving the designs of the next generation of refractive surgical laser systems. Two common refractive laser systems, variable-spot scanning (type A) and small-spot scanning (type B), are discussed by identifying sources of error that could adversely affect the capability of these lasers to accurately produce complex, customized wavefront guided ablations. A mathematical model was used to construct a laser simulator that models the two common laser systems in terms of the root-mean-square error. Error sources from ablation profile fitting, ablation registration, eye tracking, and the laser delivery system are compared, and the relative contribution of each to the overall system error is analyzed. This system-level analysis can be helpful to the improvement of both laser systems.
Keywords :
atmospheric VOC , isoprene , inhibition of S(IV) autoxidation , isoprene oxidation , Sulphur dioxide