Title of article
TLC–FID (Iatroscan) analysis of heavy oil and tar sand samples
Author/Authors
Chunqing Jiang ، نويسنده , , Steve R. Larter، نويسنده , , Kimberley J. Noke، نويسنده , , Lloyd R. Snowdon، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
5
From page
1210
To page
1214
Abstract
TLC–FID (Iatroscan) reproducibility of the fractionation [saturated hydrocarbons, aromatic hydrocarbons, resins, asphaltenes (SARA)] of heavy oil and tar sands samples has been optimized by avoiding overloading and channeling caused by very small sample spot sizes at the origin of the silica - coated rods (S-III). Doping the rods with copper (II) sulfate increased the signal response for the saturate and aromatic fractions and resulted in a decrease in the relative standard deviation of the measured peak areas for all four SARA fractions. Comparison of precipitated asphaltenes and column chromatographic separation of the maltene fraction with Iatroscan data is problematic because resins are typically occluded in the precipitated asphaltenes and polar fractions appear to have variable flame ionization detector (FID) response factors.
Journal title
Organic Geochemistry
Serial Year
2008
Journal title
Organic Geochemistry
Record number
753851
Link To Document