Title of article :
A continuous monitor-sulfur chemiluminescence detector (CM-SCD) system for the measurement of total gaseous sulfur species in air
Author/Authors :
Douglas L. MacTaggart، نويسنده , , Sherry O. Farwell، نويسنده , , Julia R. Burdge، نويسنده , , Zhong-Tao Cai، نويسنده , , Timothy J. Haakenson، نويسنده , , W. Lee Bamesberger، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
A continuous monitor-sulfur chemiluminescence detector (CM-SCD) system with a flameless, temperature-controlled furnace combustion source was developed for real-time measurement of total sulfur gases in air. This measurement system demonstrated a linear dynamic range exceeding five orders of magnitude and equimolar sensitivity to the most prevalent atmospheric sulfur gases. A detection limit of 10 pptv was obtained using 10 min signal averages. On a real-time basis, detection in the 20–50 pptv range was demonstrated. After modification of its sample inlet system, the CM-SCD showed no appreciable interference effects from the addition of H2O vapor, NO2 or O3 to the sample matrix. In the recent Gas-Phase Sulfur Intercomparison Experiment (GASIE), the CM-SCD compared favorably with SO2 measurement methods.
Keywords :
Sulfur chemiluminescence detection , Continuous monitors , Sulfur gases
Journal title :
Atmospheric Environment
Journal title :