Direct bidirectional angle-insensitive imaging of the flow signal intensity in Doppler optical coherence tomography
Author/Authors
Piao، Daqing نويسنده , , Zhu، Quing نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
-347
From page
348
To page
0
Abstract
We introduce a new method, to our knowledge, for direct detection of flow signal intensity by stationary target rejection. In our system, two delay lines are constructed with identical scanning speed and ranging depth. One delay line is used for depth ranging as well as phase modulation, and the other one acts as a full-range retroreflector (FRRR). The signal from this FRRR carries the overall features of local phase modulation, and it is used as the local oscillator for coherent demodulation. With this setup, stationary targets can be rejected at a 4-kHz high-pass cutoff frequency of the filter that follows the demodulator, compared with 20 kHz for conventional fixedfrequency demodulation. This technique features angle insensitivity and provides flow direction as well by implementing standard in-phase and quadrature detection. Besides the direct directional detection of flow signal intensity, flow speed information can be acquired with postprocessing.
Keywords
medical , Medical optics , Biotechnology , Optical coherence tomography , Measurement , Metrology , Imaging systems , instrumentation , biological imaging , Medical optics instrumentation
Journal title
Applied Optics
Serial Year
2005