Title of article :
Differences between real and predicted corneal shapes after aspherical corneal ablation
Author/Authors :
Anera، Rosario G. نويسنده , , Villa، César نويسنده , , Jiménez، José R. نويسنده , , Gutiérrez، Ram?n نويسنده , , Barco، Luis Jiménez del نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
We study the differences between real and expected corneal shapes, using an aspherical ablation algorithm with a known equation and avoiding the limitation imposed by most studies of refractive surgery in which the ablation equations are not known. We have calculated the theoretical corneal shape predicted by this algorithm, comparing this shape with the real corneal topography. The results indicate that the deviations that appear in the corneal shape are significant for visual performance and for the correction of eye aberrations. If we include in this analysis the effect of reflection losses and nonnormal incidence on the cornea, we can reduce corneal differences, but they will remain significant. These results confirm that it is essential to minimize corneal differences to achieve effective correction in refractive surgery.
Keywords :
Medical optics instrumentation , Ophthalmic optics
Journal title :
Applied Optics
Journal title :
Applied Optics