• Title of article

    Strategies for Laser-Induced Fluorescence Detection of Nitric Oxide in High-Pressure Flames. III. Comparison of A -X Excitation Schemes

  • Author/Authors

    Hanson، Ronald K. نويسنده , , Lee، Tonghun نويسنده , , Schulz، Christof نويسنده , , Jeffries، Jay B. نويسنده , , Bessler، Wolfgang G. نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -4921
  • From page
    4922
  • To page
    0
  • Abstract
    Laser-induced fluorescence (LIF) has proven a reliable technique for nitric oxide (NO) diagnostics in practical combustion systems. However, a wide variety of different excitation and detection strategies are proposed in the literature without giving clear guidelines of which strategies to use for a particular diagnostic situation. We give a brief review of the high-pressure NO LIF diagnostics literature and compare strategies for exciting selected transitions in the A-X (0, 0), (0, 1), and (0, 2) bands using a different detection bandpass. The strategies are compared in terms of NO LIF signal strength, attenuation of laser and signal light in the hot combustion gases, signal selectivity against LIF interference from O2 and CO2 , and temperature and pressure sensitivity of the LIF signal. The discussion is based on spectroscopic measurements in laminar premixed methaneair flames at pressures between 1 and 60 bars and on NO and O2 LIF spectral simulations.
  • Keywords
    harmonizable functions , covariance , Consistency , multipath , cyclostationary , Spectral density function , Bias , Doppler , Estimation , aliasing