    Comparing the methods of measuring compartment pressures in acute compartment syndrome

  • Author/Authors

    Annemarie Uliasz، نويسنده , , Jay T. Ishida، نويسنده , , Jason K. Fleming، نويسنده , , Loren G. Yamamoto، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    3
  • From page
    143
  • To page
    145
  • Abstract
    The objectives of this study were to compare the reliability and accuracy of published methods of measuring intramuscular compartment pressures. A muscle compartment model was constructed. Pressure was standardized using a water column exerting pressure on a muscle slab. The following three methods of intracompartmental pressure were used to measure the pressure within the muscle slab of the model: (1) Stryker, (2) manometric IV pump, and (3) Whitesides method. The Stryker and IV pump methods were similarly accurate compared with the “standard pressure.” The Whitesides method was not reliable in measuring a reproducible pressure. The IV pump method is comparable to the Stryker instrument in terms of accuracy and ease of use. If a Stryker instrument is unavailable, the IV pump method can serve as a suitable and satisfactory alternative.
  • Journal title
    American Journal of Emergency Medicine
  • Serial Year
    2003