Title of article :
Biphasic extrathoracic cuirass ventilation for resuscitation
Author/Authors :
Ilan Gur، نويسنده , , Ephraim Bar-Yishay، نويسنده , , Ron Ben-Abraham، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
4
From page :
488
To page :
491
Abstract :
Purposes The MRTX portable lightweight respirator (MRTX) provides noninvasive respiratory support using biphasic extrathoracic ventilation via a cuirass fitted around the patientʹs chest. Methods MRTX was applied with or without full protective gear, on adult volunteers simulating nerve agent (NA) victims by nonmedical caregivers. Assessment was made based on scores for correct positioning of the cuirass, quality of seal, and rapid ness. Results For the unprotected and protected personnel, the respective median (±95% confidence interval) scores for correct positioning of the cuirass were 2 (1.4-1.9) and 1 (1.2-1.8) (n = 15 per group, P = NS); quality of seal scores were 2 (1.5-2.0) and 2 (1.3-1.8) (P = NS); and mean (±SD) time required for instituting mechanical ventilation was 90.5 ± 10.9 and 100.3 ± 7.9 seconds (P < .05). The respirator was activated at first attempt 11 times in the group of 15 without protective gear and 8 times in the group of 15 with protective gear (P = NS). Discussion Biphasic cuirass ventilation is an easily learned and rapidly applied method suitable for use by nonmedical personnel, even when wearing cumbersome protective gear.
Journal title :
American Journal of Emergency Medicine
Serial Year :
2005
Journal title :
American Journal of Emergency Medicine
Record number :
780730
Link To Document :
بازگشت