Title of article :
Improving calibration of 3-D video oculography systems
Author/Authors :
K.، Schreiber, نويسنده , , T.، Haslwanter, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
-675
From page :
676
To page :
0
Abstract :
Eye movement recordings with video-based techniques have become very popular, as long as they are restricted to the horizontal and vertical movements of the eye. Reliable measurement of the torsional component of eye movements, which is especially important in the diagnosis and investigation of pathologies, has remained a coveted goal. One of the main reasons is unresolved technical difficulties in the analysis of video-based images of the eye. Based on simulations, we present solutions to two of the primary problems: a robust and reliable calibration of horizontal and vertical eye movement recordings, and the extraction of suitable iris patterns for the determination of the torsional eye position component.