Author/Authors :
Wang، Ge نويسنده , , J.T.، Rubinstein, نويسنده , , S.K.، Yoo, نويسنده , , F.، Collison, نويسنده , , M.W.، Vannier, نويسنده , , N.H.، Kim, نويسنده , , H.-J.، Kim, نويسنده ,
Abstract :
Three-dimensional (3-D) localization of individual cochlear implant electrodes within the inner ear is of importance for modeling the electrical field of the cochlea, designing the electrode array, and programming the associated speech processor. A 3-D reconstruction method of cochlear implant electrodes is proposed to localize individual electrodes from two X-ray views in combination with the spiral computed tomography technique. By adapting epipolar geometry to the configuration of an X-ray imaging system, we estimate individual electrode locations in the least square sense without using a patient attachment required by an existing stereophotogrammetry technique. Furthermore, our method does not require any knowledge of the intrinsic and extrinsic parameters of the imaging system. The performance of our method is studied in numerical simulation and with patient data and is found to be sufficiently accurate for clinical use. The maximum root meansquare errors measured are 0.0445 and 0.214 mm for numerical simulation and patient data, respectively.